logo.gif

THE USE OF TIME CONSTANT CIRCUITS IN ELECTROLEVEL APPLICATIONS



When an ELECTROLEVEL tilt transducer is measuring an angle which varies with time, the output voltage depends on the dynamic characteristics of both the ELECTROLEVEL sensor and the electronic detector used to drive the sensor.
For small oscillations the ELECTROLEVEL sensor behaves like a damped simple pendulum with a characteristic natural frequency fn and a damping factor .
When the ELECTROLEVEL is oscillated at frequencies above the natural frequency the apparent amplitude of the oscillations, as measured by a standard detector, falls off with increasing frequency.
Thus high frequency vibrations from machinery etc., do not affect the ELECTROLEVEL measurement unless they are of very high amplitude.
Signals at frequencies lower than the natural frequency are reproduced faithfully down to steady state or d.c. values.


ELECTROLEVEL standard sensors have the following dynamic parameters:-

VialType

Natural frequency

damping factor @ 25°C

A

1.0 Hz

0.6

C

2.5 Hz

0.9

In common with all viscous damping mechanisms these values of damping factor increase at lower temperature and decrease at higher temperature.  Natural frequency however is insensitive to temperature.
It is possible to modify the above standard damping factors to suit particular applications.
These two parameters, natural frequency and damping factor can be used in standard formulae to evaluate the ELECTROLEVEL response in a given application (Figure 1).
In many cases the great flexibility of electronic circuitry can be used to tailor the dynamic response to suit the application. In certain applications, for example when measuring the Heel angle of a ship, the steady-state angle is being measured and the oscillating component (caused in this case by the Roll motion of the ship due to wave action) is not required. .
To reduce this oscillating component a time constant circuit can be fitted to the electronic detector. A simple time constant circuit with characteristic time constant T attenuates signals with frequency above 2 /T at a rate of 6dB per octave (Figure 2).
When specifying the value of T for a given application, the settling time for a new steady reading to stabilise must be taken into account. Following a sudden step change in the steady reading, the new reading will reach 90% of its final value after 2.3T and 99% of its final value in 4.6T.
The final choice of T is thus a trade-off between settling time for the desired reading and the attenuation of the oscillating signals. Values of T in the range 2 - 60 seconds are typical.
More complex circuits can be incorporated in the electronic detector, which provide sharper cut-off at high frequencies with the optimum settling time, or to enable a variety of time constants to be switched into circuit.
Note that because the ELECTROLEVEL is a pendulous device, when it is mounted on a moving structure, it will respond to horizontal accelerations of the structure in the plane of rotation of the sensor. A separate application note The ELECTROLEVEL as an accelerometer describes this in more detail.


Frequency Response

Figure 1 Frequency Response
Relative gain plotted against frequency f
plotted normalised to natural frequency fn for
various damping factors 0.4 to 2.0

Settling Response

Figure2 Frequency Response
Relative gain (dB) plotted
against frequency f scaled
by filter time constant T


.
Tilt Measurements' Engineers are available for discussion of any possible applications and have many years experience in the design of systems for customers from a diverse range of industries.
Contact Sales for more information


All information herein is believed to be correct but no liability is accepted by Tilt Measurement Limited for any application in respect of fitness of purpose, infringement of intellectual property rights,or consequential loss or damage howsoever caused.

  Download Data Sheet as .PDF File (64Kbytes)

ã 2008 Tilt Measurement Limited